- Product Details
- Parameter
Technical Specifications
Voltage Inputs (V1, V2, V3, VN, V4, V4N)
Standard (Un) |
400 VLN/690 VLL +20% |
Range |
5 V to 2 Un for 400 VLN nominal |
Overload |
2 x Un continuous, 4 x Un for 1 s |
Burden |
< 0.5 VA/per phase |
PT Ratio |
Primary |
1-1,000,000 V |
Secondary |
1-1,500 V |
V4 Primary |
1-1,000,000 V |
V4 Secondary |
1-1,500 V |
Measurement Category |
CAT III 1000 V |
Frequency |
40 Hz-60 Hz @ 50 Hz, 48 Hz-72 Hz @ 60 Hz |
Current Inputs (·I11, I12, ·I21, I22, ·I31, I32, ·I41, I42)
Standard (In) |
5 A (Standard), 1 A (Optional) |
|
|
Range |
1% to 400% In |
Starting Current |
0.1% In |
Overload |
4 x In continuous, 10 x In for 1 s |
Burden |
< 0.5 VA/per phase @ 5 A |
< 0.1 VA/per phase @ 1 A |
CT Ratio |
Primary |
1-30,000 A |
Secondary |
1-50 A |
I4 Primary |
1-30,000 A |
I4 Secondary |
1-50 A |
SCCP Options
Split-Core Current Probe
Input @ max. 500mV |
SCCPs (50A, 200A, 500A)
Rogowski Coil (5000A max.) |
Power Supply (L/+, N/-)
Standard |
95-250VAC/VDC ±10%, 47-440 Hz |
Optional |
20-60VDC |
Burden |
< 14VA/10W @ 250VAC/DC, < 6W @ 24VDC |
Overvoltage Category |
OVC III 300V |
Digital Inputs (DIC, DI1, DI2, DI3, DI4, DIC2, DI5, DI6, DI7, DI8)
Standard |
Dry contact, 24 VDC internally wetted |
Sampling |
1000 Hz |
Hysteresis |
1 ms minimum |
Digital Outputs (DO11, DO12, DO21, DO22, DO31, DO32, DO41, DO42)
Type |
Form A Mechanical Relay |
Loading |
5A @ 250VAC/30VDC |
Alarm Output (Alarm)
Loading |
5A @ 250VAC or 30VDC |
Optional Solid State Pulse Outputs (E1+, E1-, E2+, E2-)
Type |
Form A Solid State Relay |
Isolation |
Optical |
Max. Load Voltage |
30VDC |
Max. Forward Current |
100mA |
Optional Analog Inputs (AI1+, AI1-, AI2+, AI2-, SH)
Type |
0-20/4-20 mA DC |
Overload |
24 mA maximum |
Optional Temperature Inputs (TC11, TC12, TC21, TC22, SH)
RTD Type |
2-Wire PT100 (sensor not included) |
Measurement Range |
-40°C to +200°C |